Comparison of a near-infrared reflectance spectroscopy system and skin conductance measurements for in vivo estimation of skin hydration: a clinical study
Ilona ZablockaJanis SpigulisJanis KisisInga SakniteAleksejs Zavorinssubject
lcsh:Applied optics. PhotonicsspectroscopyskinOptical fiberlcsh:Medical technologyAcoustics and UltrasonicsDiffuse reflectance infrared fourier transformwaterBiomedical EngineeringAbsorption (skin)near-infraredlaw.inventionBiomaterialsOpticsIn vivolawSpectroscopySpectrometerintegumentary systembusiness.industryChemistryNear-infrared spectroscopylcsh:TA1501-1820Atomic and Molecular Physics and Opticslcsh:R855-855.5Diffuse reflectanceDiffuse reflectionbusinessabsorptionhydrationdescription
Diffuse reflectance spectroscopy system was developed for estimation of skin hydration in the near-infrared spectral range of 900-1700 nm. Experimental setup consisted of a near-infrared spectrometer, Y-type fiber optics probe with 1 detection and 6 illumination fibers, halogen-tungsten light source and a PC. By analyzing diffuse reflectance spectrum, a parameter representing skin hydration by performing baseline correction and calculating the area under the 1450 nm water absorption maximum is proposed. A clinical study was performed acquiring data of skin hydration of 39 patients' forearm skin. Results of the developed system are compared to results obtained by a commercial device based on skin conductance measurements.
year | journal | country | edition | language |
---|---|---|---|---|
2017-04-01 | Journal of Biomedical Photonics & Engineering |