Ceftazidime Determination in Serum by High-pressure Liquid Chromatography
Till HumbertUlrich FauthAndreas Rümelinsubject
Peak areaDetection limitChromatographyChemistryReproducibility of ResultsCeftazidimeReference StandardsCeftazidimeHigh-performance liquid chromatographyCephalosporinsLinear relationshipSimple sampleDrug DiscoverymedicineHumansQuantitative analysis (chemistry)Chromatography High Pressure Liquidmedicine.drugAntibacterial agentdescription
A rapid and sensitive high-pressure liquid chromatographic method with simple sample preparation was developed for the quantitative analysis of the beta-lactam antibiotic ceftazidime (CAS 78439-06-2, Fortum). A good linear relationship was established between the peak area and the amount of ceftazidime injected over a concentration range of 1 to 200 microg/ml. The detection limit of the method was calculated to be 0.9 microg/ml. Stability was shown at 4 degrees C and at -196 degrees C for time periods of 2 h and 84 days, respectively.
year | journal | country | edition | language |
---|---|---|---|---|
2004-07-30 | Arzneimittelforschung |