Clinical efficacy of a computerised device (STA™) and a pressure syringe (VarioJect INTRA™) for intraligamentary anaesthesia.

Bilal Al-nawasPeer W. KämmererEik SchiegnitzP. KämmererMonika DaubländerB. WillershausenT. Von HaussenN. Shabazfar

subject

Periodontal LigamentAnesthesia DentalDentistryCarticaineMandibleInferior alveolar nerveEducationInjectionsstomatognathic systemMedicinePeriodontal fiberHumansClinical efficacyProspective StudiesAnesthetics LocalProspective cohort studyGeneral DentistrySyringePain MeasurementDental curriculumbusiness.industrySyringesGold standardSoft tissueAnesthesiaCase-Control StudiesClinical CompetencebusinessAnesthesia Local

description

Introduction To evaluate the impact of experience whilst using different devices for intraligamentary anaesthesia, dental students tested clinical anaesthetic efficacy of a mechanical (PDL-S) and a computer-controlled (CCLAD) application system in restorative patients. Material and methods In a prospective study, 41 Patients in need of restorative treatment in lower posterior mandible were randomised into three groups (PDL-S: teeth n = 22; CCLAD: teeth n = 20; inferior alveolar nerve block (IANB): teeth n = 20). Dental last year students conducted anaesthesia and dental treatment. Primary objectives were differences in pain during treatment [numeric rating scale (NRS)] as well as in anaesthetic efficacy (complete/sufficient vs. insufficient/no effect) between PDL-S and CCLAD. Pain of injection, need for second injections, amount of anaesthetic solution and duration of local numbness were assessed. IANB as gold standard was compared to each system descriptively. Results PDL-S had a significant lower pain during treatment (P = 0.017) but a similar anaesthetic efficacy (P = 0.175) compared to the CCLAD system. Concerning pain of injection (P = 0.42), quantity of second injections (P = 0.232), amount of used solution (P = 0.073) as well as duration of local numbness (P = 0.253), no differences were seen. When comparing both periodontal ligament injection (PDL)-systems with IANB, pain rating was higher when using CCLAD. Injection pain, amount of used solution as well as soft tissue anaesthesia was less for PDL-S and CCLAD. Discussion Both PDL techniques showed a good success for dental routine treatments. Although, compared to the mechanic device, the CCLAD system might need more clinical daily experience. Conclusion We recommend including specific trainings in intraligamentary anaesthesia in the dental curriculum.

10.1111/eje.12096https://pubmed.ncbi.nlm.nih.gov/24646115