6533b82efe1ef96bd1293bf6
RESEARCH PRODUCT
SU-FF-T-41: Monte Carlo Dosimetric Study of the New BEBIG Co-60 HDR Source
E. CasalS AgramuntDomingo GraneroR CasesFacundo BallesterJose Perez-calatayudsubject
medicine.medical_specialtyComputer sciencemedicine.medical_treatmentMonte Carlo methodBrachytherapyGeneral MedicineDose distributionHigh-Dose Rate BrachytherapyMonte carlo codemedicineDosimetryMedical physicsDose rateAlgorithmdescription
Purpose: The use of high dose rate brachytherapy (HDR) is a highly extended practice today, being the Ir‐192 the most widely extended isotope used for this type of practice although Co‐60 is also available for HDR. The purpose of this study is to obtain the dosimetric parameters of the Co‐60 source used by the BEBIG MultiSource remote afterloader (BEBIG GmbH, Germany) for which there is no dosimetric data available in the literature. It is recommended that accurate dose distribution data, based on a realistic geometry and on the mechanical characteristics of the source, should be obtained by an appropriate method, experimental or Monte Carlo, to be used as input in the HDR Treatment Planning System. Method and Materials: The Monte Carlo code GEANT4 has been used to obtain the TG43 parameters and the 2‐D rectangular dose rate table of the BEBIG Co‐60 HDR source. Results: The dose rate constant, radial dose function and anisotropy function have been calculated and are presented in tabular form as well as a detailed 2‐D rectangular dose rate table to check the treatment planning systems calculations. Conclusion: A Monte Carlodosimetric study of the BEBIG 60 Co HDR source for which no published dosimetric data exists has been done. These dosimetric datasets can be used as input and/or to validate the TPS calculations.
year | journal | country | edition | language |
---|---|---|---|---|
2005-05-26 | Medical Physics |