Cubic smoothing splines background correction in on-line liquid chromatography–Fourier transform infrared spectrometry

Miguel De La GuardiaDavid CarriónGuillermo QuintásJulia KuligowskiSalvador Garrigues

subject

Analytical chemistrySensitivity and SpecificityBiochemistryPolyethylene GlycolsAnalytical ChemistryMatrix (chemical analysis)ChemometricsSmoothing splinesymbols.namesakeSpectroscopy Fourier Transform InfraredCalibrationComputer SimulationFourier transform infrared spectroscopyAnalysis of VarianceChromatographyElutionChemistryOrganic ChemistryGreen Chemistry TechnologyGeneral MedicineFourier transformAlcoholsLinear ModelssymbolsBackground Correction MethodAlgorithmsChromatography Liquid

description

A background correction method for the on-line coupling of gradient liquid chromatography and Fourier transform infrared spectrometry (LC-FTIR) is proposed. The developed approach applies univariate background correction to each variable (i.e. each wave number) individually. Spectra measured in the region before and after each peak cluster are used as knots to model the variation of the eluent absorption intensity with time using cubic smoothing splines (CSS) functions. The new approach has been successfully tested on simulated as well as on real data sets obtained from injections of standard mixtures of polyethylene glycols with four different molecular weights in methanol:water, 2-propanol:water and ethanol:water gradients ranging from 30 to 90, 10 to 25 and from 10 to 40% (v/v) of organic modifier, respectively. Calibration lines showed high linearity with coefficients of determination higher than 0.98 and limits of detection between 0.4 and 1.4, 0.9 and 1.8, and 1.1 and 2.7 mgmL⁻¹ in methanol:water, 2-propanol:water and ethanol:water, respectively. Furthermore the method performance has been compared with a univariate background correction approach based on the use of a reference spectra matrix (UBC-RSM) to discuss the potential as well as pitfalls and drawbacks of the proposed approach. This method works without previous variable selection and provides minimal user-interaction, thus increasing drastically the feasibility of on-line coupling of gradient LC-FTIR.

https://doi.org/10.1016/j.chroma.2010.05.033