Assessment of anterior segment measurements using a high-resolution imaging device

Ramón Ruiz-mesaSalvador Aguilar-córcolesRobert Montés-micóPedro Tañá-rivero

subject

MaleMaterials sciencegenetic structuresAnterior ChamberBiomedical EngineeringIrismacromolecular substances030204 cardiovascular system & hematology03 medical and health sciences0302 clinical medicineOpticsOptical coherence tomographyLens CrystallinemedicineHumansmedicine.diagnostic_testbusiness.industryResolution (electron density)Reproducibility of ResultsGeneral MedicineRepeatabilityeye diseasesnervous systemFemaleSurgerysense organsbusinessAlgorithmsTomography Optical Coherence030217 neurology & neurosurgery

description

To assess the repeatability of several anterior segment parameters obtained with swept-source optical coherence tomography (SS-OCT).We measured 5-times consequently several parameters in 69 eyes using the ANTERION SS-OCT. White-to-white (WTW), angle-to-angle (ATA), spur-to-spur (STS), and lens vault distances were measured in the horizontal- and vertical-meridian. Anterior-chamber-angle (ACA), scleral-spur-angle (SSA), angle-opening-distance (AOD), and trabecular-iris-space-area (TISA) were measured at 500 and 750 μm for the superior-, nasal-,inferior-, and temporal-quadrants. Intrasubject standard deviation (Sw),coefficient of variation (CoV), coefficient of repeatability (CoR), and intraclass-correlation-coefficient (ICC) were calculated for each parameter. Bland-Altman analysis was done.We have not found statistically significant differences between repeated measurements (p0.05). Repeatability was good for the different parameters evaluated. Sw values in distances and areas were low and ranged from 0.01 to 0.07. CoR values showed a similar pattern being larger for those metrics measuring angles. The same happened with CoV values, being very small for WTW,ATA, and STS distances (0.16-0.57%). ICC values for all parameters analyzed were0.97. Bland-Altman plots evidenced the narrow limits of agreement for all parameters.The ANTERION SS-OCT demonstrated high repeatability measuring different distances, angles, and areas of the anterior segment of the eye.

https://doi.org/10.1080/17434440.2020.1816463