Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis : an in vitro study
Alejandro Estrada-martínezEsteban Pérez-pevidaDavid Chávarri-pradoOier Montalbán-vadilloAntonio Jiménez-garrudoIratxe Viteri-agustínMarkel Diéguez-pereiraAritza Brizuela-velascosubject
Insertion torqueMaterials scienceResearch0206 medical engineeringImplant design030206 dentistry02 engineering and technology:CIENCIAS MÉDICAS [UNESCO]020601 biomedical engineering03 medical and health sciencesResonance frequency analysis0302 clinical medicinemedicine.anatomical_structureCoronal planeUNESCO::CIENCIAS MÉDICASmedicineIn vitro studyCortical boneImplantOral SurgeryGeneral DentistryBiomedical engineeringdescription
Background This study aimed to evaluate the effect of the implant design and the presence of cortical bone in the primary stability, as well as analyze the differences between the stability measurements obtained by two different resonance frequency analysis (RFA) devices. Material and methods A total of 80 Klockner implants of two different models [40 Essential Cone implants (group A) and 40 Vega implants (group B)] were used. The implants were placed in two polyurethane blocks that simulated the mechanical properties of the maxillary bone. One block featured a layer of cortical bone that was absent from the other block. The primary stability of all implants was measured by insertion torque and RFA using two different devices: Penguin RFA and Osstell IDX. Results Primary stability was superior in the cortical bone in both torque and RFA. In the block containing cortical bone, group A implants obtained a greater insertion torque than did group B. The insertion torque was lesser in the bone lacking cortex. Regarding the ISQ of the implants, group A presented higher values in the block with cortical bone, but the values were lower in the block without cortical bone. There were no significant differences between the values obtained from the Osstell IDX and Penguin RFA. Conclusions The presence of cortical bone positively influences the primary stability of dental implants. The design of the implant also has a statistically significant influence on implant primary stability, although the impact depends on whether there is coronal cerclage or not. There were no statistically significant differences in the implant stability measurements obtained by two different devices. Key words:Implant stability, resonance frequency analysis, torque, osstell, penguin, cortical.
| year | journal | country | edition | language |
|---|---|---|---|---|
| 2020-03-01 |