6533b85afe1ef96bd12b9505
RESEARCH PRODUCT
Chromatography of metal chelates
P. HeizmannKarlheinz Ballschmitersubject
chemistry.chemical_classificationAqueous solutionChromatographyResolution (mass spectrometry)Clinical BiochemistryGeneral MedicineBiochemistryAnalytical ChemistryMetalSolventColumn chromatographychemistryPulmonary surfactantvisual_artvisual_art.visual_art_mediumGeneral Materials ScienceChelationDithiocarbamatedescription
The chromatographic behavior of metal bis(ethoxyethyl)dithiocarbamate was investigated. The chelates of 14 elements (V, Cr, Co, Ni, Cu, Zn, Se, Mo, Cd, Te, Hg, Tl, Pb and Bi) could be preconcentrated from water with an on-line system and were shown to be suitable for C18-reversed-phase liquid chromatography. A quaternary solvent mixture with admixture of a surfactant was used as eluent, UV-detection was performed at 254 nm. Thermostating the column to 40° C resulted in a significant increase in chromatographic resolution and in a decrease of analysis time. The recovery of the chelates from the aqueous solutions as well as reproducibility of the results are very sensitive to pH changes.
year | journal | country | edition | language |
---|---|---|---|---|
1973-01-01 | Fresenius' Zeitschrift f�r Analytische Chemie |