6533b85dfe1ef96bd12be707
RESEARCH PRODUCT
FT-Raman determination of Mepiquat chloride in agrochemical products
Miguel De La GuardiaGuillermo QuintásAgustín PastorSalvador Garriguessubject
Peak areachemistry.chemical_compoundChromatographyFt ramanResolution (mass spectrometry)ChemistryCalibrationAnalytical chemistryMepiquat chlorideAmmoniumScrew capFourier transform ramanSpectroscopydescription
Abstract A fast and environmentally friendly procedure has been developed for Fourier transform Raman (FT-Raman) determination of a quaternary ammonium compound, Mepiquat chloride, in pesticide emulsifiable concentrate formulations. The method is based on the measurement of peak area values between 711 and 701 cm −1 corrected using a baseline defined at 620 cm −1 and interpolated in an external calibration of Mepiquat chloride standards dissolved in water. FT-Raman spectra were collected over 1.5 min from samples confined in screw cap standard chromatographic glass vials, at a nominal resolution of 4 cm −1 accumulating 25 scans per spectra and using a laser power of 1000 mW. Five commercial samples containing around 4.5% (w/v) Mepiquat chloride (3.8% (w/v) Mepiquat) were analysed by the proposed FT-Raman procedure and results obtained were statistically comparable with those obtained by a chromatographic reference procedure. The developed FT-Raman procedure does not use toxic or harmful solvents and also minimises the contact of the operator with toxic products, being a fast alternative to the chromatographic procedure.
year | journal | country | edition | language |
---|---|---|---|---|
2004-10-01 | Vibrational Spectroscopy |