6533b85efe1ef96bd12c095e

RESEARCH PRODUCT

First-order derivative resolution of overlapped PAH peaks with common mass spectra in gas chromatography–mass spectrometry

Bernhard LendlMiguel De La GuardiaAgustín PastorGuillermo Quintás

subject

Detection limitFluoranthenechemistry.chemical_compoundChromatographyResolution (mass spectrometry)ChemistryMass spectrumAnalytical chemistryDerivativeGas chromatographyGas chromatography–mass spectrometryMass spectrometryAnalytical Chemistry

description

Abstract First-order derivative of 15 points smoothed chromatograms of mixtures of benzo( b )fluoranthene (BbF) and benzo( k )fluoranthene (BkF) for the m / z 252 ion obtained using two microscans has been proposed as a simple alternative to the resolution of overlapped chromatographic peaks of these two compounds which have identical mass spectra. The procedure provides limits of detection of 12 and 29 ng g −1 for BbF and BkF, respectively and offers the possibility to use the 250 and 248 ions to confirm the identity of these analytes.

https://doi.org/10.1016/j.talanta.2007.06.050