A rapid method for measuring intraocular lens power in vitro with a focimeter.

Amparo NaveaCristina Peris-martínezJosé María ArtigasMari Carmen García-domeneMaría Amparo Díez-ajenjo

subject

Lenses Intraocularmedicine.medical_specialtyOptics and PhotonicsMaterials sciencegenetic structuresmedicine.medical_treatmentReproducibility of ResultsIntraocular lensDiagnostic Techniques Ophthalmologicaleye diseasesSensory Systemslaw.inventionLens (optics)Cellular and Molecular NeuroscienceOphthalmologylawIntraocular lens powerOphthalmologymedicinesense organs

description

In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the limitations, the results show a good correlation.

10.1016/j.exer.2015.09.009https://pubmed.ncbi.nlm.nih.gov/26386149