6533b86cfe1ef96bd12c828f
RESEARCH PRODUCT
A rapid method for measuring intraocular lens power in vitro with a focimeter.
Amparo NaveaCristina Peris-martínezJosé María ArtigasMari Carmen García-domeneMaría Amparo Díez-ajenjosubject
Lenses Intraocularmedicine.medical_specialtyOptics and PhotonicsMaterials sciencegenetic structuresmedicine.medical_treatmentReproducibility of ResultsIntraocular lensDiagnostic Techniques Ophthalmologicaleye diseasesSensory Systemslaw.inventionLens (optics)Cellular and Molecular NeuroscienceOphthalmologylawIntraocular lens powerOphthalmologymedicinesense organsdescription
In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the limitations, the results show a good correlation.
year | journal | country | edition | language |
---|---|---|---|---|
2015-11-01 | Experimental eye research |