6533b86cfe1ef96bd12c8971
RESEARCH PRODUCT
Biomechanische Untersuchungen verschiedener Osteosynthesen an der proximalen Tibia - Vorstellung einer neuartigen meßgeometrischen Anordnung. Biomechanical Investigations of Various Osteosynthesis Procedures on the Proximal Tibia - Introduction of a New Measuring System
Dorothea MehlerM. HansenPol Maria Rommenssubject
Proximal tibiaMaterials scienceOsteosynthesisBiomedical EngineeringTorsion (mechanics)Ultrasonic sensorTibiaBiomedical engineeringdescription
In this article a new measuring system for biomechanical studies of various osteosynthesis procedures on the tibia is described. The paired tibias are osteotomized and embedded in PMMA in accordance with a randomization protocol. Testing is then carried out under nonaxial loading at 350 N, 600 N and 900 N, uniaxial bending applying a force of 12 Nm, and torsion at a force of 5 Nm in a universal pneumatic testing machine. Ultrasonic sensors pick up movement at the fracture gap under external loading. Testing of the measuring setup has been successful, and relevant results were obtained.
year | journal | country | edition | language |
---|---|---|---|---|
2003-01-01 | Biomedizinische Technik/Biomedical Engineering |