A multicenter measurement of magnetization transfer ratio in normal white matter
Vincent DoussetFrederik BarkhofGareth J. BarkerIsabelle BerryAdriana CampiDavid MillerPaul S. ToftsPaul S. ToftsWolfgang SchreiberJ. M. FranconiAchim Gasssubject
PhysicsDrug trialMagnetization Transfer MRIbusiness.industryWhite mattermedicine.anatomical_structureRepetition TimeFlip anglemedicineBrain mriRadiology Nuclear Medicine and imagingMagnetization transferNuclear medicinebusinessdescription
To assess the importance of intercenter variations when measuring magnetization transfer ratio (MTR) in the brain, six European centers measured MTR in normal white matter. MTR ranged from 9 to 51 percent units (25 sequences). The effective flip angle of the saturating pulse divided by the pulse repetition time (ENRsat degrees/msec) was a good predictor of MTR (MTR = 3.25 ENRsat).
year | journal | country | edition | language |
---|---|---|---|---|
1999-03-01 | Journal of Magnetic Resonance Imaging |