A new high precision voltage divider has been built for monitoring the analysing potential of the KATRIN main spectrometer. In addition a condensed 83mKr calibration source has been set up at the modified 1 eV resolving Mainz spectrometer, our measurements show that energy calibration and monitoring for KATRIN is feasible at the few ppm level.