Search results for "Oversizing"
showing 2 items of 2 documents
An In Vitro Phantom Study on the Role of the Bird-Beak Configuration in Endograft Infolding in the Aortic Arch.
2015
Purpose: To assess endograft infolding for excessive bird-beak configurations in the aortic arch in relation to hemodynamic variables by quantifying device displacement and rotation of oversized stent-grafts deployed in a phantom model. Methods: A patient-specific, compliant, phantom pulsatile flow model was reconstructed from a patient who presented with collapse of a Gore TAG thoracic endoprosthesis. Device infolding was measured under different flow and pressure conditions for 3 protrusion extensions (13, 19, and 24 mm) of the bird-beak configuration resulting from 2 TAG endografts with oversizing of 11% and 45%, respectively. Results: The bird-beak configuration with the greatest protr…
Prognostic value of the ratio between prothesis area and indexed annulus area measured by multiSlice-CT for transcatheter aortic valve implantation p…
2016
Background Postprocedural aortic regurgitations following transcatheter aortic valve implantation (TAVI) procedures remain an is- sue. Benefit of oversizing strategies to prevent them isn’t well established. We compared different level of oversizing in our cohort of con- secutive patients to address if severe oversizing compared to normal sizing had an impact on post-procedural outcomes. Methods From January 2010 to August 2013, consecutive patients were referred for TAVI with preoperative Multislice-CT (MSCT) and the procedures were achieved using Edwards Sapien® or Corevalve devices®. Retrospectively, according to pre-procedural MSCT and the valve size, pa- tients were classified into thr…