C. Candela-juan

On the use of the absorbed depth-dose measurements in the beam calibration of a surface electronic high-dose-rate brachytherapy unit, a Monte Carlo-based study.

PURPOSE To evaluate the use of the absorbed depth-dose as a surrogate of the half-value layer in the calibration of a high-dose-rate electronic brachytherapy (eBT) equipment. The effect of the manufacturing tolerances and the absorbed depth-dose measurement uncertainties in the calibration process are also addressed. METHODS The eBT system Esteya® (Elekta Brachytherapy, Veenendaal, The Netherlands) has been chosen as a proof-of-concept to illustrate the feasibility of the proposed method, using its 10 mm diameter applicator. Two calibration protocols recommended by the AAPM (TG-61) and the IAEA (TRS-398) for low-energy photon beams were evaluated. The required Monte Carlo (MC) simulations w…

research product

Backscatter and Transmission through a Lead Shield Used in Surface and Interstitial HDR Brachytherapy

research product

Calculated organ doses using Monte Carlo simulations in a reference male phantom undergoing HDR brachytherapy applied to localized prostate carcinoma

Purpose : The aim of this study was to obtain equivalent doses in radiosensitive organs (aside from the bladder and rectum) when applying high-dose-rate (HDR) brachytherapy to a localized prostate carcinoma using60Co or 192 Ir sources. These data are compared with results in a water phantom and with expected values in an infinite water medium. A comparison with reported values from proton therapy and intensity-modulated radiation therapy (IMRT) is also provided. Methods : Monte Carlo simulations in Geant4 were performed using a voxelized phantom described in International Commission on Radiological Protection (ICRP) Publication 110, which reproduces masses and shapes from an adult reference…

research product

SU-E-J-64: Evaluation of a Commercial EPID-Based in Vivo Dosimetric System in the Presence of Lung Tissue Heterogeneity

Purpose: To study the performance of Dosimetry Check (DC), an EPID-based dosimetry software, which allows performing transit dosimetry, in low density medium, by comparing calculations in-phantom, and analysing results for 15 lung patients. Methods: DC software (v.3.8, pencil beam-based algorithm) has been tested, for plans (Eclipse v.10.0 TPS) delivered in two Varian Clinac iX equipped with aS1000 EPIDs.In the CIRS lung phantom, comparisons between DC and Eclipse (Acuros) were performed for several plans: (1) four field box; (2) square field delivered in arc mode; (3) RapidArc lung patient plan medially centred; (4) RapidArc lung patient plan centred in one lung. Reference points analysed:…

research product

Frequency of dicentrics and contamination levels in Ukrainian children and adolescents from areas near Chernobyl 20 years after the nuclear plant accident.

International audience; Purpose To survey the possible presence of chromosomal damage and internal contamination in a group of Ukrainian children and adolescents, 20 years after the Chernobyl accident at the Nuclear Power Plant. Materials and methods Cytogenetical procedures were performed according to dicentric assay in 55 Ukrainian children and adolescents (29 boys and 26 girls), living near Chernobyl. In addition, a whole body detector and urinalysis were used to detect internal contamination. Results 36 dicentrics were found in a total of 53,477 metaphases scored in these children, which reflected a frequency of dicentrics below the background level. On the other hand, internal contamin…

research product

SU-E-T-516: Limitations and Clinical Implications of the TG-43 Formalism for High-Dose-Rate Skin Brachytherapy

Purpose: In high‐dose‐rate (HDR) skin brachytherapy, sources are located outside, in contact with, or implanted at some depth below the skin surface. Nowadays, most TPS use the TG‐43 formalism that is based on sources within an infinite water medium, without accounting for the scatter defect due to surrounding air. The purpose of the present study is to evaluate the limitations of the TG‐43 formalism in HDR skin brachytherapy. Methods: This study examined two HDR sources (Ir‐192 and Co‐60) together with a hypothetical Yb‐169 source. Dose distributions were obtained using Monte Carlo methods (i.e., GEANT4). The following configurations were investigated: 1)A typical clinical case consisting …

research product

EP-1590: Shielding design and fetal dose evaluation of a breast cancer pregnant patient undergoing HDR Ir-192 brachytherapy

research product

New HDR Valencia Applicator for Treating Skin Lesions Larger Than 3 cm Size with Either a Co-60 or Ir-192 Source

research product

SU-F-T-63: Dosimetric Relevance of the Valencia and Leipzig HDR Applicators Plastic Cap

Purpose: Utilization of HDR brachytherapy treatment of skin lesions using collimated applicators, such as the Valencia or Leipzig is increasing. These applicators are made of cup-shaped tungsten material in order to focalize the radiation into the lesion and to protect nearby tissues. These applicators have an attachable plastic cap that removes secondary electrons generated in the applicator and flattens the treatment surface. The purpose of this study is to examine the dosimetric impact of this cap, and the effect if the cap is not placed during the HDR fraction delivery. Methods: Monte Carlo simulations have been done using the code Geant4 for the Valencia and Leipzig applicators. Dose r…

research product

SU-E-T-508: Monte Carlo Dosimetric Study of the CSM40 Low Dose Rate Source

Purpose: There is a LDR 137Cs brachytherapy source, CSM40, developed by Eckert & Ziegler BEBIG. Fundamental dosimetric datasets to be used on treatment planning systems do not exist for this source. This study aims to obtain quality dosimetric data for this source for its use in clinical practice as required by AAPM and ESTRO. Methods: Penelope2008 and GEANT4 Monte Carlo codes were used to characterize this source dosimetry. Water composition and mass density was that recommended by AAPM. Due to the typical energies of the 137 Cs, dose for small distances can be approximated by collisional kerma. Therefore, we have considered collisional kerma for r<20 cm. To provide adequate spatial resolu…

research product

Influence of the simultaneous calibration of multiple ring dosimeters on the individual absorbed dose

Abstract Ring dosimeters for personal dosimetry are calibrated in accredited laboratories following ISO 4037-3 guidelines. The simultaneous irradiation of multiple dosimeters would save time, but has to be carefully studied, since the scattering conditions could change and influence the absorbed dose in nearby dosimeters. Monte Carlo simulations using PENELOPE-2014 were performed to explore the need to increase the uncertainty of H p 0 .07 in the simultaneous irradiation of three and five DXT-RAD 707H-2 (Thermo Scientific) ring dosimeters with beam qualities: N-30, N-80 and N-300. Results show that the absorbed dose in each dosimeter is compatible with each of the others and with the refere…

research product

Design and characterization of a new high-dose-rate brachytherapy Valencia applicator for larger skin lesions

Purpose: The aims of this study were (i) to design a new high-dose-rate (HDR) brachytherapy applicator for treating surface lesions with planning target volumes larger than 3 cm in diameter and up to 5 cm in size, using the microSelectron-HDR or Flexitron afterloader (Elekta Brachytherapy) with a 192Ir source; (ii) to calculate by means of the Monte Carlo(MC) method the dose distribution for the new applicator when it is placed against a water phantom; and (iii) to validate experimentally the dose distributions in water. Methods: The penelope2008MC code was used to optimize dwell positions and dwell times. Next, the dose distribution in a water phantom and the leakage dose distribution arou…

research product

WE-C-108-08: Organ Doses in a Male Phantom Undergoing High-Dose-Rate Brachytherapy Applied to the Prostate

Purpose: The aim of this study was to obtain equivalent doses to radiosensitive organs when applying high‐dose‐rate (HDR) brachytherapy to the prostate using60 Co or 192 Ir sources, and in comparison to external‐beam radiotherapy (EBRT) modalities. Methods: Monte Carlo simulations in Geant4 were performed using a voxelized adult reference man described in Publication 110 by the International Commission on Radiological Protection (ICRP). Point sources of 60Co or 192Ir with photon energy spectra corresponding to those exiting their capsules were placed in the center of the prostate. Equivalent doses per therapeutic absorbed dose to the prostate were obtained in several radiosensitive organs. …

research product

Dosimetric perturbations of a lead shield for surface and interstitial high-dose-rate brachytherapy.

In surface and interstitial high-dose-rate brachytherapy with either (60)Co, (192)Ir, or (169)Yb sources, some radiosensitive organs near the surface may be exposed to high absorbed doses. This may be reduced by covering the implants with a lead shield on the body surface, which results in dosimetric perturbations. Monte Carlo simulations in Geant4 were performed for the three radionuclides placed at a single dwell position. Four different shield thicknesses (0, 3, 6, and 10 mm) and three different source depths (0, 5, and 10 mm) in water were considered, with the lead shield placed at the phantom surface. Backscatter dose enhancement and transmission data were obtained for the lead shields…

research product

SU-E-J-69: Evaluation of the Lens Dose On the Cone Beam IGRT Procedures

Purpose: With the establishment of the IGRT as a standard technique, the extra dose that is given to the patients should be taken into account. Furthermore, it has been a recent decrease of the dose threshold in the lens, reduced to 0.5 Gy (ICRP ref 4825-3093-1464 on 21st April, 2011).The purpose of this work was to evaluate the extra dose that the lens is receive due to the Cone-Beam (CBCT) location systems in Head-and-Neck treatments. Methods: The On-Board Imaging (OBI) v 1.5 of the two Varian accelerators, one Clinac iX and one True Beam, were used to obtain the dose that this OBI version give to the lens in the Head-and-Neck location treatments. All CBCT scans were acquired with the Sta…

research product

Response to “Comment on ‘Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry’ ” [Med. Phys. 42 , 4954-4964 (2015)]

research product

Dosimetric characterization of two radium sources for retrospective dosimetry studies

Purpose: During the first part of the 20th century, Ra-226 was the most used radionuclide for brachytherapy. Retrospective accurate dosimetry, coupled with patient follow up, is important for advan ...

research product

Evaluation of the shielding in a treatment room with an electronic brachytherapy unit.

Esteya® (Elekta Brachytherapy, Veenendaal, The Netherlands) is an electronic brachytherapy (eBT) system based on a 69.5 kVp x-ray source and a set of collimators of 1 to 3 cm in diameter, used for treating non-melanoma skin cancer lesions. This study aims to estimate room shielding requirements for this unit. The non-primary (scattered and leakage) ambient dose equivalent rates were measured with a Berthold LB-133 monitor (Berthold Technologies, Bad Wildbad, Germany). The latter ranges from 17 mSv h-1 at 0.25 m distance from the x-ray source to 0.1 mSv h-1 at 2.5 m. The necessary room shielding was then estimated following US and some European guidelines. The room shielding for all barriers…

research product

Calibration of a thermoluminescent dosimeter worn over lead aprons in fluoroscopy guided procedures

Fluoroscopy guided interventional procedures provide remarkable benefits to patients. However, medical staff working near the scattered radiation field may be exposed to high cumulative equivalent doses, thus requiring shielding devices such as lead aprons and thyroid collars. In this situation, it remains an acceptable practice to derive equivalent doses to the eye lenses or other unprotected soft tissues with a dosimeter placed above these protective devices. Nevertheless, the radiation backscattered by the lead shield differs from that generated during dosimeter calibration with a water phantom. In this study, a passive personal thermoluminescent dosimeter (TLD) was modelled by means of …

research product

Commissioning and periodic tests of the Esteya® electronic brachytherapy system

A new electronic brachytherapy unit from Elekta, called Esteya(®), has recently been introduced to the market. As a part of the standards in radiation oncology, an acceptance testing and commissioning must be performed prior to treatment of the first patient. In addition, a quality assurance program should be implemented. A complete commissioning and periodic testing of the Esteya(®) device using the American Association of Physicists in Medicine (AAPM), Groupe Europeen de Curietherapie and the European Society for Radiotherapy & Oncology (GEC-ESTRO) guidelines for linacs and brachytherapy units as well as our personal experience is described in this paper. In addition to the methodology, r…

research product

Evaluation of lens absorbed dose with Cone Beam IGRT procedures

The purpose of this work is to evaluate the absorbed dose to the eye lenses due to the cone beam computed tomography (CBCT) system used to accurately position the patient during head-and-neck image guided procedures. The on-board imaging (OBI) systems (v.1.5) of Clinac iX and TrueBeam (Varian) accelerators were used to evaluate the imparted dose to the eye lenses and some additional points of the head. All CBCT scans were acquired with the Standard-Dose Head protocol from Varian. Doses were measured using thermoluminescence dosimeters (TLDs) placed in an anthropomorphic phantom. TLDs were calibrated at the beam quality used to reduce their energy dependence. Average dose to the lens due to …

research product

Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

Purpose: A surface electronic brachytherapy (EBT) device is in fact an x-ray source collimated with specific applicators. Low-energy (<100 kVp) x-ray beam dosimetry faces several challenges that need to be addressed. A number of calibration protocols have been published for x-ray beam dosimetry. The media in which measurements are performed are the fundamental difference between them. The aim of this study was to evaluate the surface dose rate of a low-energy x-ray source with small field applicators using different calibration standards and different small-volume ionization chambers, comparing the values and uncertainties of each methodology. Methods: The surface dose rate of the EBT unit …

research product

Fetal dose measurements and shielding efficiency assessment in a custom setup of (192)Ir brachytherapy for a pregnant woman with breast cancer.

To assess the radiation dose to the fetus of a pregnant patient undergoing high-dose-rate (HDR) (192)Ir interstitial breast brachytherapy, and to design a new patient setup and lead shielding technique that minimizes the fetal dose.Radiochromic films were placed between the slices of an anthropomorphic phantom modeling the patient. The pregnant woman was seated in a chair with the breast over a table and inside a leaded box. Dose variation as a function of distance from the implant volume as well as dose homogeneity within a representative slice of the fetal position was evaluated without and with shielding.With shielding, the peripheral dose after a complete treatment ranged from 50 cGy at…

research product

SU-E-T-343: Valencia Applicator Commissioning Using a Micro-Chamber Array

Purpose: In the commissioning and QA of surface isotope-based applicators, source-indexer distance (SID) has a great influence in the flatness, symmetry and output. To these purposes, methods described in the literature are the use of a special insert at the entrance of dwell chamber or radiochromic films. Here we present the experience with a micro-chamber array to perform the commissioning and QA of Valencia applicators. Methods: Valencia applicators have been used, the classic and the new extra-shielded version. A micro-chamber array has been employed, 1000 SRS (PTW), with 977 liquid filled, 2.3×2.3×0.5 mm3 sized ion chambers covering 11×11 cm2, which spacing is 2.5 mm in the central 5.5…

research product

SU-E-T-691: Shielding Evaluation of a Treatment Room with An Electronic Brachytherapy System

Purpose: Esteya™ (Elekta) is a brachytherapy electronic system used for treating skin cancer lesions. This unit is based on a 69.5 kV X-ray source and a surface applicator which produces a circular beam of a specific diameter, which varies from 1 to 3 cm. This study aims to establish the radiation protection measures for the system. Methods: A characterization of the scattered and leakage radiation of the system was implemented by means of experimental measurements. The scattered radiation measurements were performed with a Berthold LB133 ionization chamber, evaluating the dose rate for different distances from the applicator surface. The patient was simulated through water equivalent solid…

research product

Dosimetric Deviation of Leipzig and Valencia Skin Applicators When Used without Cap

research product

Assaying multiple 125 I seeds with the well-ionization chamber SourceCheck 4π 33005 and a new insert

PURPOSE To provide a practical solution that can be adopted in clinical routine to fulfill the AAPM-ESTRO recommendations regarding quality assurance of seeds used in prostate permanent brachytherapy. The aim is to design a new insert for the well-ionization chamber SourceCheck(4π) 33005 (PTW, Germany) that allows evaluating the mean air-kerma strength of up to ten (125)I seeds with one single measurement instead of measuring each seed individually. MATERIAL AND METHODS The material required is: a) the SourceCheck(4π) 33005 well-ionization chamber provided with a PTW insert to measure the air-kerma strength S K of one single seed at a time; b) a newly designed insert that accommodates ten s…

research product

Commissioning and quality assurance procedures for the HDR Valencia skin applicators

The Valencia applicators (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) are cup-shaped tungsten applicators with a flattening filter used to collimate the radiation produced by a high-dose-rate (HDR) 192 Ir source, and provide a homogeneous absorbed dose at a given depth. This beam quality provides a good option for the treatment of skin lesions at shallow depth (3-4 mm). The user must perform commissioning and periodic testing of these applicators to guarantee the proper and safe delivery of the intended absorbed dose, as recommended in the standards in radiation oncology. In this study, based on AAPM and GEC-ESTRO guidelines for brachytherapy units and our experience, a set …

research product