6533b85bfe1ef96bd12bac04
RESEARCH PRODUCT
Evaluation of the shielding in a treatment room with an electronic brachytherapy unit.
Jose Perez-calatayudJ.a. Bautista-ballesterosC. Candela-juanJavier VijandeJavier VijandeBlanca Ibanez-roselloJ. I. VillaescusaFacundo Ballestersubject
0106 biological sciencesbusiness.industryEquivalent dosemedicine.medical_treatmentBrachytherapyBrachytherapyPublic Health Environmental and Occupational HealthTreatment roomGeneral MedicineRadiation Dosage01 natural sciences030218 nuclear medicine & medical imaging03 medical and health sciences0302 clinical medicineRadiation ProtectionElectromagnetic shieldingCalibrationmedicineHumansScattering RadiationNuclear medicinebusinessRadiometryWaste Management and Disposal010606 plant biology & botanydescription
Esteya® (Elekta Brachytherapy, Veenendaal, The Netherlands) is an electronic brachytherapy (eBT) system based on a 69.5 kVp x-ray source and a set of collimators of 1 to 3 cm in diameter, used for treating non-melanoma skin cancer lesions. This study aims to estimate room shielding requirements for this unit. The non-primary (scattered and leakage) ambient dose equivalent rates were measured with a Berthold LB-133 monitor (Berthold Technologies, Bad Wildbad, Germany). The latter ranges from 17 mSv h-1 at 0.25 m distance from the x-ray source to 0.1 mSv h-1 at 2.5 m. The necessary room shielding was then estimated following US and some European guidelines. The room shielding for all barriers considered was below 2 mmPb. The dose to a companion who, exceptionally, would stay with the patient during all treatment was estimated to be below 1 mSv if a leaded apron is used. In conclusion, Esteya shielding requirements are minimal.
year | journal | country | edition | language |
---|---|---|---|---|
2017-03-02 | Journal of radiological protection : official journal of the Society for Radiological Protection |