0000000000388612

AUTHOR

J. Rosello Ferrando

SU-E-I-83: Detectability Limits of a New Positron Emission Mammography in Relation to Tumour-Size, Tumour-To-Background Ratio and Activity Concentration

Purpose:Positron emission mammography(PEM) improves spatial resolution and sensitivity, making it suitable for early breast tumours detection. The aim of this study is to evaluate the limits of a dedicated breast PET in terms of tumour‐size, tumour‐to‐background activity concentration ratio (TBR) and activity concentration. Methods: A dedicated PEM is evaluated. To characterize the device, we use a phantom of 15 cm of diameter containing 6 inserts of inner diameters 18, 15, 11, 8, 5 and 3 mm. To evaluate the detectability limits images are acquired by varying the TBR from 10 to 2 and without background activity concentration. In all the studies the activity concentration for the 6 inserts i…

research product

Characterization of alpha sources by Rutherford backscattering spectrometry

Radioactive sources for alpha spectrometry are usually prepared by electrodeposition onto stainless steel backings (and sometimes heated). In earlier work, using the conventional method with passivated implanted planar silicon detectors for the measurements, several sources had been characterized in terms of various parameters by fitting the data of each spectrum to a certain mathematical function. In the present work, the Rutherford Backscattering Spectrometry (RBS) technique with a 1.6 MeV He+ beam was used to study the influence of those factors on the surface distribution and depth profiles of the thin radionuclide layers. Simulations of the measurements using the RUMP computer code wer…

research product