Thermal lens spectrometric determination of cerium with oxine
Abstract The spectrophotometric and photothermal (TLS) procedures for cerium determination using 8-hydroxyquinoline (oxine), after extraction into chloroform, are compared. Photothermal measurements are made using a coaxial pump/probe thermal lens spectrometer. The use of high-purity reagents at low concentrations permits a decrease in the TLS blank signal and noise, leading to a limit of detection of 9 × 10−9 M (cerium extract concentration), 40-fold lower than the spectrophotometric value. The dynamic range extended up to 6 × 10−7 M and the relative standard deviation for 5 × 10−7 M cerium was 3.9%.
Determination of catecholamines in urine by micellar liquid chromatography with coulometric detection
The determination of catecholamines by HPLC with a sodium dodecyl-sulphate (SDS), micellar mobile phase on a C18 column and with coulometric detection was studied. The eluate was conditioned at +0.25 and +0.00 V, and the current at −0.16V was recorded. A previously developed model which describes the chromatographic behaviour of solutes in HPLC with hybrid, micellar mobile phases was used to optimize the SDS and ethanol concentrations. A mobile phase of 0.15M SDS in a phosphate buffer of pH 3.4 and without ethanol is recommended. The limits of detection were 0.4–0.7 ng ml−1. The procedure was applied to the determination of unconjugated L-dopa, norepinephrine and dopamine in urine. Direct i…