Development of silicon pad detectors and readout electronics for a Compton camera
Abstract Applications in nuclear medicine and bio-medical engineering may profit using a Compton camera for imaging distributions of radio-isotope labelled tracers in organs and tissues. These applications require detection of photons using thick position-sensitive silicon sensors with the highest possible energy and good spatial resolution. In this paper, research and development on silicon pad sensors and associated readout electronics for a Compton camera are presented. First results with low-noise, self-triggering VATAGP ASIC's are reported. The measured energy resolution was 1.1 keV FWHM at room temperature for the 241 Am photo-peak at 59.5 keV .
First coincidences in pre-clinical Compton camera prototype for medical imaging
Abstract Compton collimated imaging may improve the detection of gamma rays emitted by radioisotopes used in single photon emission computed tomography (SPECT). We present a crude prototype consisting of a single 500 μm thick, 256 pad silicon detector with pad size of 1.4×1.4 mm 2 , combined with a 15×15×1 cm 3 NaI scintillator crystal coupled to a set of 20 photo multipliers. Emphasis is placed on the performance of the silicon detector and the associated read-out electronics, which has so far proved to be the most challenging part of the set-up. Results were obtained using the VATAGP3, 128 channel low-noise self-triggering ASIC as the silicon detector's front-end. The noise distribution (…