A 20-channel receive-only mouse array coil for a 3 T clinical MRI system
A 20-channel phased-array coil for MRI of mice has been designed, constructed, and validated with bench measurements and high-resolution accelerated imaging. The technical challenges of designing a small, high density array have been overcome using individual small-diameter coil elements arranged on a cylinder in a hexagonal overlapping design with adjacent low impedance preamplifiers to further decouple the array elements. Signal-to-noise ratio (SNR) and noise amplification in accelerated imaging were simulated and quantitatively evaluated in phantoms and in vivo mouse images. Comparison between the 20-channel mouse array and a length-matched quadrature driven small animal birdcage coil sh…
Design and evaluation of a 32-channel phased-array coil for lung imaging with hyperpolarized 3-helium
Imaging with hyperpolarized 3-helium is becoming an increasingly important technique for MRI diagnostics of the lung but is hampered by long breath holds (>20 sec), which are not always applicable in patients with severe lung disease like chronic obstructive pulmonary disease (COPD) or α-1-anti-trypsin deficiency. Additionally, oxygen-induced depolarization decay during the long breath holds complicates interpretation of functional data such as apparent diffusion coefficients. To address these issues, we describe and validate a 1.5-T, 32-channel array coil for accelerated 3He lung imaging and demonstrate its ability to speed up imaging 3He. A signal-to-noise ratio increase of up to a factor…