An harmonic test condition for the metrological characterization and the operating principle identification of reactive static meters

Valentina CosentinoAntonio CataliottiSalvatore NuccioAlessandro Lipari

subject

EngineeringTotal harmonic distortionbusiness.industryAcousticsAstrophysics::Instrumentation and Methods for AstrophysicsAC powerMetrologyElectric power systemHarmonicsHarmonicElectronic engineeringMetreenergy measurements harmonic distortion power quality power systemsbusinessSettore ING-INF/07 - Misure Elettriche E ElettronicheVoltage

description

This paper faces the problem of the metrological characterization of the static meters for reactive energy in the presence of harmonic distortion. Starting from the only available accuracy test condition in the presence of harmonics introduced in the Standards for active static meters, an experimental approach is proposed for the verification of such meters in the presence of harmonic distortion, as well as for the individuation of their operating principle. A theoretical validation of the obtained results is presented by means of a comparison with a standard and a commercial static meter with a known operating principle.

https://doi.org/10.1109/imtc.2008.4547237