Influence of short implants geometry on primary stability.

José González-serranoRicardo OrtegaJuan López-quilesLuis-miguel Sáez-alcaideBeatriz Pardal-peláezPedro Molinero-mourelle

subject

0206 medical engineeringDentistry02 engineering and technologyBone and BonesDental Prosthesis Retention03 medical and health sciences0302 clinical medicineHounsfield scaleBone qualityMedicineAnimalsGeneral Dentistrybusiness.industryResearchDental Implantation Endosseousfood and beverages030206 dentistry:CIENCIAS MÉDICAS [UNESCO]020601 biomedical engineeringShort implantsOtorhinolaryngologyDental Prosthesis DesignUNESCO::CIENCIAS MÉDICASSurgeryCattleImplantOral Surgerybusiness

description

Background A correct design is needed in short implants to improve primary stability (PS) in low quality bone. This study aimed to compare PS of double thread and single thread short implants. Material and Methods Thirty implants with single thread design (PHI/SHORT-I) and 30 implants with double thread design (PHIA/SHORT-I) (Radhex®, Inmet-Garnick S.A., Guadalajara, Spain) were placed in 30 randomly selected bovine ribs. PS was assessed in implant stability quotients (ISQ) and periotest values (PV) with Osstell™ and Periotest® devices, respectively. Computed tomographies of the ribs were taken and bone quality was evaluated in Hounsfield Units (HU) using Ez3D Plus software (Vatech Co., Korea). Only implants placed in low quality bone according to Misch and Kircos classification were selected (D3 bone: 350-850 HU; and D4 bone: 150-350 HU). Ten implants were not included in the study for being placed in D1 and D2 bone. Finally, 50 implants were selected: 17 and 9 PHI/SHORT-I in D3 and D4 bone respectively, and 15 and 9 PHIA/SHORT-I in D3 and D4 bone respectively. Results The one-way ANOVA showed statistically significant differences in ISQ (61.35 ± 4.77 in PHI/SHORT-I and 66.43 ± 4.49 in PHIA/SHORT-I, P<0.005) and PV (-2.76 ± 0.8 and -4.11 ± 1.24 respectively, P<0.005) between two implant designs in D3 bone, and statistically significant differences in ISQ (53.44 ± 3.34 in PHI/SHORT-I and 60.56 ± 1.53 in PHIA/SHORT-I, P<0.0001) and PV (1.13 ± 0.95 and -2.5 ± 0.61 respectively, P<0.0001) between two groups in D4 bone. Conclusion Double thread design short implants resulted to have higher PS in comparison with single thread design short implants in D3 and D4 bone. Key words:ISQ, osstell, periotest, primary stability, resonance frequency analysis, short implants.

https://hdl.handle.net/20.500.12530/43693