6533b82ffe1ef96bd129463b
RESEARCH PRODUCT
A wire-loop technique for implantation of an iliac branched device in a patient with previous surgery for a ruptured abdominal aortic aneurysm
Gaetano VitaleUmberto BracaleDonatella NareseGuido BajardiFranco MaglionePietro Orlando Contisubject
Malemedicine.medical_specialtyRadiology Nuclear Medicine and ImagingBridging (networking)Common iliac aneurysmProsthesis DesignIliac ArterySettore MED/22 - Chirurgia VascolareWire-loop techniqueFollow-Up StudieBlood Vessel Prosthesis ImplantationmedicineStentHumanscardiovascular diseasesAgedEndovascular; Iliac aneurysm; Wire-loop technique; Aged; Aortic Aneurysm Abdominal; Blood Vessel Prosthesis Implantation; Follow-Up Studies; Humans; Iliac Aneurysm; Iliac Artery; Male; Prosthesis Design; Treatment Outcome; Blood Vessel Prosthesis; Stents; Radiology Nuclear Medicine and Imaging; Cardiology and Cardiovascular MedicineRuptured abdominal aortic aneurysmEndovascularbusiness.industrySynthetic graftStandard techniqueSurgeryBlood Vessel ProsthesisIliac aneurysmBlood Vessel Prosthesisurgical procedures operativemedicine.anatomical_structureTreatment OutcomeStentsRadiologybusinessCardiology and Cardiovascular MedicineWire loopArteryFollow-Up StudiesAortic Aneurysm AbdominalHumandescription
We described a modified technique for implanting a bridging stent-graft into an iliac branched device. A 79-year-old male who had received aortobiiliac synthetic graft surgery for a ruptured abdominal aortic aneurysm six months earlier was admitted to our unit for treatment of a left common iliac aneurysm involving the origin of the hypogastric artery. A standard technique was unsuccessful at implanting the bridging stent-graft, and therefore a wire-loop guidewire over the graft bifurcation was used to stabilize the contralateral sheaths and to complete the implantation. © Turkish Society of Radiology 2012.
year | journal | country | edition | language |
---|---|---|---|---|
2013-01-01 |