6533b833fe1ef96bd129b5f6
RESEARCH PRODUCT
Transition-edge microcalorimeters for X-ray space applications
Heikki SipiläK. M. KinnunenJukka P. PekolaAntti NuottajärviArttu Luukanensubject
PhysicsSuperconductivitybusiness.industryBolometerX-rayMicrobolometerCondensed Matter PhysicsSpace (mathematics)Electronic Optical and Magnetic MaterialsCalorimeterlaw.inventionSQUIDOpticslawThermometerOptoelectronicsElectrical and Electronic Engineeringbusinessdescription
Abstract In an European Space Agency funded research project, our goal is to develop microbolometer technology for X-ray and far-infrared detection for ESA's future scientific missions. We report results on the X-ray calorimeter, which is based on the superconducting transition of the Ti/Au thermometer strip at about 200 mK. Incident X-rays heat up a Bi absorber, deposited on top of the 400 μm ×400 μm thermometer. The temperature rise of the absorber is measured as a change of the thermometer current with a SQUID operating at 1 K.
year | journal | country | edition | language |
---|---|---|---|---|
2000-07-01 | Physica B: Condensed Matter |