6533b836fe1ef96bd12a1312

RESEARCH PRODUCT

Low noise scintillation detectors with a P-47 thin layer screen for electrons of several keV

P. GriesbachK. HildebrandM. A. C. LigtenbergR. AlbrechtW. MeiselR. KurzZs. KajcsosP. GütlichCh. Sauer

subject

PhysicsNuclear and High Energy PhysicsPhotomultiplierScintillationRange (particle radiation)Physics::Instrumentation and Detectorsbusiness.industryDetectorElectronScintillatorPhotoelectric effectKinetic energyOpticsbusinessInstrumentation

description

Abstract The applicability of a low-noise scintillation detector (ScD) for the registration of electrons of several keV energy has been studied employing photomultipliers (PM) of different types and sizes. With the application of a sedimented P-47 scintillation screen, the values of the low-energy sensitivity limit and those of the light conversion coefficient were determined as about 2.7–4.7 keV and 2.8–6.6 photoelectrons/keV, respectively, for the set of PM's (Philips-Valvo XP 2020, Philips-Valvo XP 2052, Philips-Valvo XP 2972, EMI 9124a) studied. It is concluded that such scintillation detectors might be used advantageously as electron counters in the range of E > 5 keV. Applications below this kinetic energy value are also feasible when applying a floating acceleration of several kV to the ScD — a voltage much lower than the values required for Everhart-Thornley detectors.

https://doi.org/10.1016/0168-583x(94)95641-3