Determination of caffeine in tea samples by Fourier transform infrared spectrometry

Guillermo QuintásSalvador GarriguesMiguel De La GuardiaJanina Ohnsmann

subject

Detection limitChromatographyTeaChemistryExtraction (chemistry)Infrared spectroscopyBiochemistryAnalytical ChemistryPlant LeavesAbsorbancechemistry.chemical_compoundColumn chromatographyCaffeineSpectroscopy Fourier Transform InfraredFourier transform infrared spectroscopyDiethyl etherCaffeine

description

A sustainable and environmentally friendly procedure has been developed for the FTIR determination of caffeine in tea leaf samples. The method is based on the extraction with ammonia and CHCl3 and direct determination of caffeine on the chloroform extracts using peak height absorbance measurements at 1658.5 cm(-1) and external calibration. The method provides a sensitivity of 0.2142 absorbance units mg(-1) mL and a limit of detection of 1 mg L(-1), corresponding to 0.002% m/m caffeine in tea leaves. As compared with a reference procedure, based on UV absorbance measurement at 276 nm after low pressure column chromatography, the developed procedure reduces the consume of CHCl3 by a factor of 10, that of NH4OH by a factor of 20 and avoids the use of diethyl ether and Celite. The time required to do the analysis of a sample is 15 minutes as compared with the 6 hours for the reference one.