6533b855fe1ef96bd12b0b0f
RESEARCH PRODUCT
Time of flight measurements based on FPGA and SiPMs for PET–MR
Albert AguilarJose M. BenllochJose TorresAntonio GonzálezRaimundo Garcia-olcinaJulio MartosFilomeno SanchezJ. SoretPedro A. Martinezsubject
PhysicsNuclear and High Energy Physicsbusiness.industryDetectorSilicon photomultipliersConstant fraction discriminatorPositron emission tomography and magnetic resonanceSignalTime-to-digital converterTime of flightTime of flightOpticsSilicon photomultiplierField Programable Gate ArrayTime-to-digital converterbusinessField-programmable gate arrayInstrumentationElectronic circuitdescription
Coincidence time measurements with SiPMs have shown to be suitable for PET/MR systems. The present study is based on 3 x 3 mm(2) SiPMs, LSO crystals and a conditioning signal electronic circuit. A Constant Fraction Discriminator (CFD) is used to digitalize the signals and a TDC FPGA-implemented is employed for fine time measurements. TDC capability allows processing the arrival of multiple events simultaneously, measuring times under 100 ps. The complete set-up for time measurements results on a resolution of 892 +/- 41 ps for a pair of detectors. The details of such implementation are exposed and the trade-offs of each configuration are discussed. (C) 2013 Elsevier By, All rights reserved,
year | journal | country | edition | language |
---|---|---|---|---|
2014-01-11 | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |