6533b858fe1ef96bd12b614c

RESEARCH PRODUCT

A high resolution electrostatic time-of-flight spectrometer with adiabatic magnetic collimation

Ernst W. OttenL. BornscheinJ. BonnCh. WeinheimerB. Degen

subject

PhysicsNuclear and High Energy PhysicsTime of flightLuminosity (scattering theory)SpectrometerMeasuring instrumentContext (language use)ElectronAtomic physicsAdiabatic processInstrumentationCharged particle

description

Abstract A new type of spectrometer for low energy charged particles is presented. It consists of an adiabatic magnetic collimation and two filters: an electrostatic retarding potential to set a lower limit (high pass) and a time-of-flight analysis to reject high energy charged particles (low pass). Both filters are only limited in their resolution by the efficiency of the adiabatic magnetic collimation. The proof of this principle is demonstrated by a pilot measurement on the K conversion line of 83mKr. Possible applications to pulsed and continuous electron sources are discussed with the emphasis on the investigation of the β spectrum of T2 to deduce information on the mass of the electron antineutrino and possible anomalies in the β spectrum. In this context design parameters of a spectrometer with a resolving power of E/ Δ E=20 000 and a luminosity of A Δ Ω/4 π =4 cm 2 for 20 keV electrons are given.

https://doi.org/10.1016/s0168-9002(98)01263-7