6533b860fe1ef96bd12c391a
RESEARCH PRODUCT
Preparation methods to optimize the performance of sensor discs for fast chemiluminescence ozone analyzers.
Ivonne TrebsAlexander MoravekFranz X. MeixnerM. BeckG. SongJens-christopher MayerR. OswaldM. Ermelsubject
OzoneLuminescent MeasurementsChemistryAnalytical chemistryGeneral Chemistrylaw.inventionPreparation methodMatrix (chemical analysis)chemistry.chemical_compoundAdsorptionOzoneEnergy TransferlawReagentLuminescent MeasurementsEnvironmental ChemistrySensitivity (control systems)Coloring AgentsChemiluminescencedescription
Fast ozone (O(3)) measurements (1-50 Hz) in the atmosphere are required for airborne studies and for the measurement of ground-based O(3) fluxes by the eddy covariance technique. Fast response analyzers, based on heterogeneous chemiluminescence, need dye coated sensor discs on which the chemiluminescence is generated. In this study, we present three new preparation methods for those sensor discs. Currently available sensor discs exhibit a fast temporal decay of sensitivity, resulting in short duty times which is troublesome for many field applications. To produce sensor discs that provide more stable signals over time, three dyes and nine energy transfer reagents were tested (as well as different stoichiometric mixtures). The resulting optimal method saves 80% of the solid chemicals and shows a duty ozone dose that is prolonged by a factor of 3.5, revealing the same average sensitivity as currently available discs. In addition, we observed a strong effect of the adsorption matrix on the O(3) sensitivity, although silica discs from the same manufacturer were used. Application of the new sensor discs during field measurements showed that the results are consistent with the laboratory data.
year | journal | country | edition | language |
---|---|---|---|---|
2013-02-01 | Environmental sciencetechnology |