6533b872fe1ef96bd12d30b1
RESEARCH PRODUCT
Pacemaker therapy in premature children with high degree AV block.
Friedrich WippermannChristoph KampmannEwald HimmrichOliver PrzibilleJürgen MeyerFranz-xaver SchmidBernd Nowaksubject
Epicardial leadMalePacemaker ArtificialHigh amplitudebusiness.industryEpicardial pacingInfant NewbornGeneral MedicineTwins MonozygoticPacemaker pulse generatorElectrodes ImplantedHeart BlockAnesthesiaBlock (telecommunications)MedicineHumansCardiology and Cardiovascular MedicinebusinessInfant Prematuredescription
The smallest pacemaker pulse generator and a steroid-eluting bipolar epicardial lead were implanted in two premature children with symptomatic AV block. Stable capture threshold and high amplitude evoked response electrogram resulted in normal function of the pacemaker Autocapture algorithm, which adjusts output 0.3 V above the measured capture threshold. Autocapture had previously been used only with endocardial leads. Longer-term observation is required.
year | journal | country | edition | language |
---|---|---|---|---|
1998-12-01 | Pacing and clinical electrophysiology : PACE |