6533b7d1fe1ef96bd125d777

RESEARCH PRODUCT

Quality Control of Spirometry in the Elderly

Gaetano MelilloDario OlivieriVittorio GrassiRiccardo PistelliFranco RengoRaffaele Antonelli-incalziVincenzo BelliaFilippo Catalano

subject

Pulmonary and Respiratory MedicineSpirometryCOPDReproducibilitymedicine.medical_specialtymedicine.diagnostic_testbusiness.industryWalk distanceCritical Care and Intensive Care Medicinemedicine.diseaselaw.inventionFEV1/FVC ratiolawmedicinePhysical therapybusinessCognitive impairmentSpirometerAsthma

description

We evaluated the outcome of the spirometry quality control program of the SA.R.A. multicenter project, the aim of which is the multidimensional assessment of asthma and COPD in the elderly ( ⩾ 65 yr). The factors determining this quality were also evaluated. The program was based on standardized procedures (ATS recommendations), performed by specifically trained and certified personnel; a fully-computerized spirometer with customized software was used for spirometry. A reference center made monthly controls. Overall, 638 cases and 984 controls were examined. Spirometric measurements were obtained in 607 cases and 912 controls; 508 and 747 tests with at least three acceptable curves were obtained in cases and in controls, respectively (NS). The percentage of reproducible tests ranged between 95.8% for FEV1 in controls and 87.6% for FVC in cases. The average reproducibility for FEV1 was 61.6 ml in cases and 58.3 ml in controls (NS). Cognitive impairment, shorter 6-min walk distance, and lower educational le...

https://doi.org/10.1164/ajrccm.161.4.9810093