6533b823fe1ef96bd127e20f
RESEARCH PRODUCT
A biomechanical evaluation of methods of distal humerus fracture fixation using locking compression plates versus conventional reconstruction plates.
J. KornerErich SchneiderChristoph JostenGerd DiederichsMichael ArzdorfF. Helmut LillBerend Linkesubject
medicine.medical_specialtyHumeral Fracturesmedicine.medical_treatmentOsteotomyFracture Fixation InternalFracture fixationBone plateMaterials TestingmedicineHumansOrthopedics and Sports MedicineHumerusAgedBone mineralOrthodonticsAged 80 and overOsteosynthesisbusiness.industryBiomechanicsStiffnessGeneral MedicineEquipment DesignMiddle AgedSurgeryBiomechanical Phenomenamedicine.anatomical_structureSurgerymedicine.symptombusinessBone Platesdescription
OBJECTIVES To examine the biomechanical behavior of 2 techniques of double-plate osteosynthesis for fractures of the adult distal humerus using conventional reconstruction plates and locking compression plates. DESIGN Basic science study. SETTING Experimental in vitro study. PATIENTS/PARTICIPANTS Forty fresh-frozen human distal humeri specimens. INTERVENTION Four matched groups with 10 humeri each, median age 74 years (46-95), were created using similar bone mineral density values. Two standard configurations of double-plate osteosynthesis (dorsal or 90 degrees configuration) with either conventional reconstruction plates or locking compression plates were studied for biomechanical properties of the constructs. A fracture model with a 5-mm supracondylar osteotomy gap simulating metaphyseal comminution (AO type 13-A3.3) was used. MAIN OUTCOME MEASUREMENT Stiffness testing of the constructs in anterior/posterior bending, torsion, and axial compression loading. Evaluation of alterations of the bone-implant interface and failure patterns under cyclic loading and strength testing. RESULTS The study demonstrates that primary stiffness in anterior/posterior bending and torsional loading is significantly increased by using locking compression plates in a 90 degrees configuration (P < 0.05) as compared with dorsally applied plates. The differences between the different plate types are insignificant if applied in the same configuration. It is demonstrated that none of the tested implants failed under cyclic loading within the number of cycles expected for 3 months of use. The bone-implant interface is less likely to fail during strength testing with locking compression plates. CONCLUSION The biomechanical behavior of the osteosynthesis depends more on plate configuration than plate type. Advantages of locking compression plates are only significant if compared with dorsal plate application techniques. Nevertheless, locking compression plates are helpful supplementary tools for achieving primary stable fracture fixation. This might be of considerable clinical relevance in patients with diminished bone mineral quality or in the presence of metaphyseal comminution.
year | journal | country | edition | language |
---|---|---|---|---|
2004-04-24 | Journal of orthopaedic trauma |