Cyclic stressMaterials scienceRoot canalNiTi alloyendodontic instruments02 engineering and technologyCurvatureOperative Dentistry and EndodonticsRadius of curvature (optics)03 medical and health sciencesReciprocating motion0302 clinical medicinemedicinecyclic fatigue; endodontic instruments; heat treatment; NiTi alloy; reciprocating motion; dentistry (all)reciprocating motionGeneral DentistryOrthodonticsdentistry (all)heat treatmentResearch030206 dentistryMechanical resistancecyclic fatigue021001 nanoscience & nanotechnology:CIENCIAS MÉDICAS [UNESCO]medicine.anatomical_structureNickel titaniumUNESCO::CIENCIAS MÉDICASFracture (geology)0210 nano-technology

description

Background To evaluate the cyclic fatigue resistance of two different Nickel-Titanium instruments, Reziflow (Komet, Brasseler GmbH & Co., Lemgo, Germany) and WaveOne Gold (Dentsply Maillefer, Ballaigues, Switzerland). Material and Methods Two groups of 20 different NiTi endodontic instruments of identical tip size of 0.25 mm were tested; Reziflow and Wave-One Gold primary. Cyclic fatigue testing was performed in a stainless steel simulated root canal manufactured by reproducing the instrument’s size and taper. A simulated root canal with a 90 degrees angle of curvature and 5 mm radius of curvature was constructed for the instruments tested. The centre of the curvature was 5 mm from the tip of the instrument and the curved segment of the canal was approximately 5 mm in length. Both the instruments were used in the same preset program specific for the WaveOne instruments. Each instrument was rotated until fracture occurred and the time to fracture (TtF) and the length of the fractured fragment were recorded. Means and standard deviations of TtF and fragment length were calculated and data were subjected to statical analysis (P0.05) in the mean length of the fractured fragments between the instruments. Conclusions Rezifllow instruments were associated with a significantly higher cyclic fatigue resistance than WaveOne Gold instruments. Key words:Endodontic instruments, NiTi alloy, Reciprocating motion, Cyclic Fatigue, Heat treatment.

https://hdl.handle.net/10550/70632