6533b85ffe1ef96bd12c283f
RESEARCH PRODUCT
JYFLTRAP: a Penning trap for precision mass spectroscopy and isobaric purification
Christine WeberPasi KarvonenUlrike HagerHeikki PenttiläJani HakalaSami Rinta-antilaSaidur RahamanJ. SzerypoS. KopeckyTommi EronenViki-veikko ElomaaAri JokinenJuha ÄYstöJuho RissanenAntti SaastamoinenDmitry GorelovIain MooreAnu KankainenVeli Kolhinensubject
Physicsnuclear spectroscopyNuclear and High Energy Physicsaccelerator-based physicsCyclotronPenning trapMass spectrometrykiihdytinpohjainen fysiikkaAtomic massIonlaw.inventionNuclear physicsydinrakenneDecay energylawnuclear structureydinspektroskopiaNuclear fusionPhysics::Atomic PhysicsSpectroscopyNuclear Experimentydinfysiikkadescription
In this article a comprehensive description and performance of the double Penning-trap setup JYFLTRAP will be detailed. The setup is designed for atomic mass measurements of both radioactive and stable ions and additionally serves as a very high-resolution mass separator. The setup is coupled to the IGISOL facility at the accelerator laboratory of the University of Jyväskylä. The trap has been online since 2003 and it was shut down in the summer of 2010 for relocation to the upgraded IGISOL facility. Numerous atomic mass and decay energy measurements have been performed using the time-of-flight ion-cyclotron resonance technique. The trap has also been used in several decay spectroscopy experiments as a high-resolution mass filter. peerReviewed
year | journal | country | edition | language |
---|---|---|---|---|
2012-04-01 |