6533b872fe1ef96bd12d3875

RESEARCH PRODUCT

Determination of ultratrace bismuth in milk samples by atomic fluorescence spectrometry.

Miguel De La GuardiaAgustín PastorPatricia Cava-montesinosM. Luisa Cervera

subject

Quality ControlAnalytical chemistrychemistry.chemical_elementBorohydridesMass spectrometryAtomic fluorescence spectrometrySensitivity and SpecificityFluorescenceAnalytical ChemistryBismuthCow milkDigestion (alchemy)Environmental ChemistryAnimalsArgonMicrowavesPharmacologyDetection limitHydrideSpectrum AnalysisMilkAshingchemistrySpainIndicators and ReagentsAgronomy and Crop ScienceBismuthFood Science

description

Abstract A sensitive procedure was developed for determination of bismuth (Bi) in milk samples by hydride generation atomic fluorescence spectrometry (HG–AFS) after microwave-assisted sample digestion with HNO3 and H2O2. The method provides a sensitivity of 1832 fluorescence units (ng/mL) with a detection limit of 0.01 ng/mL, which corresponds to 20 pg absolute limit of detection, equivalent to 0.50 ng/g in the original sample. Application of the methodology to cow milk samples from the Spanish market showed the presence of Bi at a concentration of 11.8–28.8 ng/g, which compared well with data obtained after dry ashing of samples and with data obtained by inductively coupled plasma–mass spectrometry after microwave-assisted digestion.

https://pubmed.ncbi.nlm.nih.gov/14509442