Influence of trifocal intraocular lenses on standard autorefraction and aberrometer-based autorefraction.

César Albarrán-diegoEsther López-arteroNuria GarzónFrancisco PoyalesMaría García-montero

subject

MaleCorneal Wavefront AberrationPseudophakiaMesopic visionZernike polynomialsmedicine.medical_treatmentVisual AcuityIntraocular lensRefraction OcularPupillaw.inventionsymbols.namesakeLens Implantation IntraocularlawmedicineHumansProspective StudiesRetinoscopyMathematicsAgedPhacoemulsificationmedicine.diagnostic_testKeratometerAberrometryMiddle AgedMultifocal Intraocular LensesSensory SystemsOphthalmologyAutorefractorsymbolsOptometrySurgeryFemalePhotopic vision

description

To study the agreement between manifest refraction and objective refraction measured with two autorefractor models and an aberrometer in eyes implanted with a trifocal diffractive intraocular lens (IOL).IOA Madrid Innova Ocular, Madrid, Spain.Prospective comparative cohort study.An autorefractor keratometer (KR-8800), based on a Scheiner double pinhole, and a 3-dimension wavefront topography aberrometer system (OPD-Scan III), based on the scanning-slit retinoscopy principle, were used to obtain objective refraction readings. In addition, lower-order Zernike coefficients (Z) were used to calculate objective refraction. A set of 7 different results was obtained in power vector notation (spherical equivalent [SE], Jackson cross-cylinder, axes at 180 degrees and 90 degrees [J0] and Jackson cross-cylinder, axes at 45 degrees and 135 degrees [J45]) for 7 different methods: manifest refraction, autorefraction obtained with the autorefractor keratometer, WF-P (Z-based objective refraction for the photopic pupil), WF-M (Z-based objective refraction for the mesopic pupil), WF-4 (Z-based objective refraction for a 4.0 mm pupil), OPD-C (autorefraction measured with the 3-dimension wavefront topography aberrometer system under photopic conditions), and OPD-M (autorefraction measured with the 3-dimension wavefront topography aberrometer system under mesopic conditions).The study comprised 102 eyes from 51 cataract patients who underwent binocular implantation of a diffractive trifocal IOL (FineVision POD F). All 6 objective methods yielded more negative SE values than manifest refraction (P  .001). As for the astigmatism components (J0 and J45), only autorefraction (P = .003) and OPD-M (P  .001) differed significantly from manifest refraction. The best and worst correlation for the SE component were intraclass correlation coefficient (ICC) = 0.70 (for WF-M) and ICC = 0.48 (for WF-4).Objective methods tend to yield more negative sphere values than manifest refraction.

10.1016/j.jcrs.2019.04.017https://pubmed.ncbi.nlm.nih.gov/31326229